OKOndt GROUP goes on operating for your non-destructive testing needs. Our team works for you to offer the best NDT solutions of top quality, and impeccable services. In September, the group of our experts attended InnoTrans 2024 and could showcase our latest developments to the visitors, as well as discuss current NDT issues and future partnership

 It is intended for determination of main parameters of ultrasonic welded joints testing carried out by various ultrasonic flaw detectors. The article provides the information on the main technical specifications and KОU-2М application areas.KОU-2М set of calibration blocks according to GOST 14782-86 It is intended for determination of main parameters of ultrasonic welded joints testing carried out by various ultrasonic flaw detectors.

KОU-2М set of calibration blocks

 It allows to perform:

  1. Selection and comparison of angle beam probes by ultrasonic waves frequency with the error of ±10% in the range of 1,25-5 МHz.
  2. Measurement of probe angle with the error of ±1°.
  3. Determination of ultrasonic beam emergent point position with the error of ±0,5 mm.
  4. Evaluation of dead area size.
  5. Check of depth gauge operation accuracy of flaw detector with the error of ±10% by SО1 and ±8% by SО2.
  6. Evaluation of equivalent defect area.
  7. Evaluation of beam resolution of flaw detector when operating straight beam and angle beam probes.
  8. Evaluation of wedge angle of angle beam probe with the error of ±2,5°.
  9. Determination of conditional sensitivity and equipment setup for the given calculated value of ultimate sensitivity with the error of ±1 mm.

Стандартные образцы КОУ-2М